Appendix B: Performance Verification
B 24
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Setup. Set the oscilloscope and AFG320 Function Generator as indicated below:
H Oscilloscope
Displayed channel
Vertical axis
. . . . .
Horizontal axis
. . .
Record length
. . . .
Acquire mode
. . . .
Acquire sequence
Trigger mode
. . . .
Trigger level
. . . . .
Trigger source
. . . .
Trigger slope
. . . . .
Trigger position
Input coupling
. . . .
Input impedance
InstaVu
. . . . . . . . .
H Function generator
Output channel
. . .
Waveform
. . . . . . .
Frequency
. . . . . . .
Amplitude
. . . . . . .
Offset
. . . . . . . . . .
Procedure. Follow the steps below to verify that the maximum operating
frequency (trigger operation) is functioning correctly.
1. Load the TP5TRIG.PDA test pattern file from the performance check disk.
2. Press the START/STOP button on the front panel. The LED, corresponding
to the button, should be on.
3. Verify that a stable data pattern as shown in Figure B-12 appears on the
oscilloscope screen.
4. Change the oscilloscope trigger source to CH4 and trigger slope to Rising
Edge.
5. Verify that a stable data pattern as shown in Figure B-13 appears on the
oscilloscope screen.
6. Press the START/STOP button on the front panel to stop output (the buttons
LED goes off).
CH1 and CH2
1 V/div for CH1 and CH2
5 V/div for CH3 and CH4
5.00 ns/div
15000
Peak Detect
.
RUN/STOP button only
Auto
2 V
CH3
Falling Edge
. .
50 %
DC
50 W for CH1 and CH2
. .
1 MW for CH3 and CH4
Infinite Persistence
CH1
Square wave
500 Hz
1 V (50 W termination)
500 mV
DG2040 Data Generator