Tektronix DG2040 User Manual page 203

Data generator
Hide thumbs Also See for DG2040:
Table of Contents

Advertisement

Appendix B: Performance Verification
Sequence & Data Output
B 10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table B 4: Internal clock frequency accuracy
Internal clock frequencies
1.100000 GHz
650.1000 MHz
200.0000 kHz
100.0000 mHz (10.0 s)
6. Press the START/STOP button on the front panel to stop the output (the
buttons LED goes off).
This test verifies that the pattern data that is output from the DG2040 Data
Generator output channel ports (CH0, CH0, CH1, and CH1) is functioning
correctly.
The following equipment is required.
Equipment Required
Connections. Refer to Figure B-4 for connections.
Use the SMA cables and the SMA female-to-BNC male adapters to make the
following connections:
H Connect CH0 output from the DG2040 Data Generator front panel to the
oscilloscope CH1 input.
H Connect CH0 output from the DG2040 Data Generator front panel to the
oscilloscope CH2 input.
Use the BNC cables to make the following connections:
H Connect the function generator CH1 output to the DG2040 Data Generator
rear panel event input.
H Connect the event output from the DG2040 Data Generator rear panel to the
oscilloscope CH3 input.
Clock frequency ranges
1.0999989 GHz  1.1000011 GHz
650.09935 MHz  650.10065 MHz
199.99980 kHz  200.00020 kHz
99.99990 mHz  100.00010 mHz
Oscilloscope
Function generator
Three BNC cables
Two SMA cables
Two SMA female to BNC male adapters
Performance check disk
DG2040 Data Generator

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dg203059021-1059021-15

Table of Contents