Edge Control Mode
Operation
DG2040 Data Generator
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This test verifies the edge control mode operation is functioning correctly.
The following equipment is required.
Equipment Required
Connections. Refer to Figure B-15 for connections.
Use the SMA cables and the SMA 5X attenuators to make the following
connections:
H Connect CH0 output from the DG2040 Data Generator front panel to the
sampling oscilloscope CH1 input.
H Connect CH1 output to the DG2040 Data Generator front panel to the
sampling oscilloscope CH2 input.
H Connect the CH1 output from the DG2040 Data Generator front panel to the
sampling oscilloscope TRIGGER INPUTS DIRECT.
H Use the remaining SMA cable, the SMA female-to-BNC male adapter, and
the 50 W termination to terminate the DG2040 Data Generator CH0 output.
Use the BNC cable to make the following connection:
H Connect the function generator CH1 output to the DG2040 Data Generator
front panel edge control input.
Appendix B: Performance Verification
Sampling oscilloscope with head
Function generator
BNC cable
Four SMA cables
SMA female to BNC male adapter
50 W termination
Three SMA X5 attenuators
Performance check disk
B 29