Appendix B: Performance Verification
AFG320
DG2040 Data Generator
CH0
CH0
Figure B 15: Edge control mode operation connection
B 30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
CH1
BNC Cable
Edge
Control
CH1
Input
CH1
SMA Cable
SMA female to BNC male adapter
and 50 W termination
Setup. Set the sampling oscilloscope and AFG320 Function Generator as
indicated below:
H Sampling oscilloscope
Displayed channel
Vertical axis
Horizontal axis
Trigger mode
Trigger level
Trigger source
Trigger slope
Trigger position
Vectored Trace
SMA 5X attenuator
11801C
CH1
CH2
TRIGGER
INPUTS
DIRECT
CH1 and CH2
. . . . .
100 mV/div for CH1 and CH2
. . .
500 ps/div
. . . .
Norm
. . . . .
0 V
. . . .
External Direct
. . . . .
Falling Edge
. .
50 %
. . .
Off
SMA 5X attenuator
DG2040 Data Generator