Tektronix DG2040 User Manual page 202

Data generator
Hide thumbs Also See for DG2040:
Table of Contents

Advertisement

DG2040 Data Generator
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
b. Set the counter trigger to an appropriate value and verify that the counter
displays a frequency in the range of 1.09999989 to 1.1000011 GHz.
Refer to Table B-4.
3. Perform the following steps to check the 650.1000 MHz clock frequency
accuracy:
a. Press the following buttons to set the DG2040 Data Generator clock
frequency to 650.1 MHz.
Menu button
Bottom button
SETUP
Clock
b. Set the counter trigger to an appropriate value, and verify that the
counter displays a frequency in the 650.09935 MHz to 650.10065 MHz
range.
4. Perform the following steps to check the 200.0000 kHz clock frequency
accuracy:
a. Change the frequency counter input to INPUT A through the 50 W
termination with the SMA cable.
b. Set the DG2040 Data Generator clock frequency to 200 kHz.
c. Set the counter trigger to an appropriate value, and verify that the
counter displays a frequency in the 199.99980 kHz to 200.00020 kHz
range.
5. Perform the following steps to check the 100.0000 mHz clock frequency
accuracy:
a. Set the DG2040 Data Generator clock frequency to 100 mHz (0.1 Hz).
b. Set the counter trigger to an appropriate value, and verify that the
counter displays a frequency in the range 99.99990 mHz to 100.00010
.
mHz
Appendix B: Performance Verification
Pop up menu
Side button
Int FREQ
Front panel
button
650.1
MHz/ms
B 9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dg203059021-1059021-15

Table of Contents