7.3 Inspect the hoses, pipelines and connections

1.
Inspect all of the electrical connections and check that they are secure. Tighten any
loose connections.
2.
Inspect all of the electrical cables and check that they are not damaged and have
not overheated. Replace or repair any damaged or overheated cable.
3.
Inspect all of the vacuum and exhaust connections and check that they are secure.
Tighten any loose connections.
4.
Inspect all of the vacuum and exhaust pipelines and check that they are not
corroded or damaged and that they do not leak. Replace or repair any corroded or
damaged component and seal any leaks found.

7.4 Factory default settings

Procedure to return the pumping system to the factory default settings.
1.
Remove the mains power from the pumping system.
2.
Press the NEXT key and hold it down whilst reapplying mains power. Continue
holding down the NEXT key until "Err01" is displayed.
3.
Remove and reapply mains power. The factory defaults will now be set.
D39595880_A
D39595880_A - Maintenance
Page 35
10/2020 - ©Edwards Limited