Figure 4.
Use of Gases
CAUTION Before using gases, carefully read the hazard indications and information reported in the
Material Safety Data Sheet (MSDS) supplied by the manufacturer referring to the CAS (Chemical
Abstract Service) number.
It is the your responsibility to see that all local safety regulations for the use of gases are obeyed.
Gas Supply
The Flash IRMS Elemental Analyzer uses the following gases:
• Helium (He) as carrier gas and purge gas [CAS number 7440-59-7]
• Oxygen (O
Note ConFlo II/III/IV and IRMS Delta models (TF Bremen manufacturing) use the following
gases:
• Helium (He) as carrier gas and purge gas [CAS number 7440-59-7]
• Carbon Monoxide (CO) as standard gas [CAS number 630-08-0]
• Hydrogen (H
• Nitrogen (N
• Carbon Dioxide (CO
• Sulfur Dioxide (SO
Gas Purity
The EA IsoLink IRMS System for CNSOH requires the use of ultra-high purity gas chromatography
grade (99.999%) gases.
Maximum Pressure
The maximum pressure of the Elemental Analyzer gas supplies is 700 kPa (7 bar).
Thermo Scientific
EA IsoLink IRMS System for CNSOH: Hot Surface Labels
) as gas for sample oxidation [CAS number 7782-44-7]
2
) as standard gas [CAS number 1333-74-0]
2
) as standard gas [CAS number 7727-37-9]
2
) as standard gas [CAS number 124-38-9]
2
) as standard gas [CAS number 7446-09-5]
2
1
Getting Familiar with your EA IsoLink IRMS System for CNSOH
Hot Surface Label
EA IsoLink IRMS System for CNSOH Operating Manual
Use of Gases
7