Starting A Sulfur Measurement; Ea Isolink Setting - Thermo Scientific EA IsoLink IRMS System Operating Manual

Including the flash irms elemental analyzer
Table of Contents

Advertisement

6
Using EA IsoLink IRMS System for CNSOH

Starting a Sulfur Measurement

22. Fill in the correct masses (64 and 66 replace for example 44, 45 and 46) to the appropriate cups
specific for your IRMS.
When highlighting the specific gas configuration by a click on its row, the number of cups required
for measurement is displayed together with the assigned masses.
23. Select a Calibration, which is valid for the selected cups.
24. Press the Save & Close button
Note
Starting a Sulfur Measurement

EA IsoLink Setting

See
Table
Table 24. Setting for EA IsoLink
Carrier
Oxygen
Reference
Cycle run time
Sampling delay
Oxygen Injection End
114
EA IsoLink IRMS System for CNSOH Operating Manual
A value of -1 denotes unlimited
24.
.
180 mL/min
25 mL/min
20 mL/min
1000 s
10 s
3 s (or higher, depending on sample size and sample type)
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EA IsoLink IRMS System and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF