Post-Analysis Operations; Putting The Instrument In Standby Mode - Thermo Scientific EA IsoLink IRMS System Operating Manual

Including the flash irms elemental analyzer
Table of Contents

Advertisement

7
Maintenance

Post-Analysis Operations

Post-Analysis Operations
For maintaining high analytical performance and to reduce operating costs, we recommend you follow
these practical suggestions.

"Putting the Instrument in Standby Mode"

"Maintaining the Instrument"
"Maintaining the Instrument"
Putting the Instrument in Standby Mode
When the work session is over, the instrument can be placed into Standby mode. This mode is
intended to reduce the gas consumption significantly and will reduce operating costs. The standby
conditions can be defined by the operator.
There are these options:
• Reduce the flow only — This option reduces the carrier and reference flow to 10 mL/min and the
oxygen to 0 mL/min.
• Define an EA method — This option allows to define settings in an EA method which are
uploaded to the EA.
The Standby function can be activated immediately or automatically at the end of the analytical
sequence.
To set the EA in standby immediately
1. Right-click on the EA window in Isodat Software Suite and open EA Standby options. When
clicking this button it opens the following page. See
Figure 131. Standby option for EA
Here you decide which standby properties you want to use when setting the EA in standby.
• Reduced gas flows (default)
120
EA IsoLink IRMS System for CNSOH Operating Manual
on
page 120
on
page 122
on
page 122
Figure
131.
Thermo Scientific