Figure 88. NC Method: Time Events List
Note By double-clicking on the relevant gas name in the time events list in column Switch Gas,
you access to Isodat Object Method Switcher window. See
Figure 89. NC Method: Acquisition and Time
Figure 90. NC Method: Method Switcher
Note Waiting Time specifies the maximum period of time, for how long the system/software will
wait for the previous peak (for example N
again, and delay the subsequent peak jump.
For example: if you enter a waiting time of 10 seconds, the software will delay the peak jump as
specified in the time event list by a maximum of 10 seconds.
• If the N
then the peak jump will occur straight away at that specified time.
• If the N
entry for waiting time is for example 10 seconds (you can also enter a different waiting time or
0 seconds here), the peak jump will be delayed by a maximum of that number of seconds, so
that the N
• If the N
then occur. If the N
will definitely occur after the waiting time period.
Thermo Scientific
peak has already reached background until the time specified in the time events list,
2
peak hasn't reached background until the specified time in the time events list and the
2
peak still has that number of seconds to reach background.
2
peak reaches background within these seconds of waiting time, the peak jump will
2
peak does not reach background within the waiting time, the peak jump
2
6
Using EA IsoLink IRMS System for CNSOH
Creating an Isodat Method for N+C Measurement (Dual Measurement)
Figure
in a dual N-C measurement), to reach background
2
EA IsoLink IRMS System for CNSOH Operating Manual
90.
89