4. Note the four sub-folders named Temp 1, Temp 2, Temp 3, and Temp 4. Each of them
corresponds to the relevant instrument (Analyzer #1, Analyzer #2, and so on), that may be
configured during the EagerSmart Data Handling Software installation procedure.
5. Double-click the sub-folder name that corresponds to your EA IsoLink IRMS System for CNSOH
for opening the relevant contents. For example: Temp 1 = Analyzer #1. Double-click on the
Eager.ini file for opening the file. See
Figure 159. Eager.ini File (1)
6. Search Pyrolysis=0 line. Indicates the configuration of EagerSmart Data Handling Software to
control a standard EA IsoLink IRMS System for CNSOH. For controlling a EA IsoLink IRMS
System for CNSOH, modify the Pyrolysis=0 line to Pyrolysis=1. See
Figure 160. Eager.ini File (2)
7. Save the modification and close the file. Run EagerSmart Data Handling Software. A message
appears indicating that the instrument is now configured to operate under pyrolysis condition.
Thermo Scientific
9
Running the Flash IRMS as Stand-alone Instrument
Configuring EagerSmart Data Handling Software
Figure
159.
EA IsoLink IRMS System for CNSOH Operating Manual
Figure
160.
151