18. Type No. If you would type Yes, this would automatically mark the template (i.e. CO
SO
2
a. Mark SO
b. If Ratio Groups other than SO
19. Confirm by OK.
20. The new Gas Configuration SO
Figure 125. Creating a New Gas Configuration
21. In the Calibration column select your current calibration file.
Note
i.e. universal triple collector. If you have a special cup configuration, the respective masses will be
collected in other cups.
Thermo Scientific
in the Ratio Groups window below.
.
2
Figure 125
shows a common cup configuration as used in most Delta mass spectrometers,
are marked, clear them all.
2
appears in the list as a row of its own.
2
EA IsoLink IRMS System for CNSOH Operating Manual
6
Using EA IsoLink IRMS System for CNSOH
Create a Gas Configuration for a Sulfur Measurement
) instead of
2
113