6
Using EA IsoLink IRMS System for CNSOH
Creating an EA Method for H+O Measurement
Note The Start Detection time must be adjusted so that the spikes on the mass traces from the
magnet jump are excluded. This may vary depending on the Time Event List setup.
23. Set the Peak Detection@CO parameters for your new dual measurement method as shown in
Figure
24. Select the Printout@CO tab of your new dual measurement method, see
Figure 114. Printout@CO Page
25. Select your Printout templates.
26. Click Save | Save as to save this method with the file ending *.met.30.
27. Enter a name for your method and click Save.
Creating an EA Method for H+O Measurement
The following figures show how an EA-method for a dual analysis H+O may be set up.
Figure 115. HO EA Method: Instrument Temperature
Note The temperature of the high-temperature conversion furnace (left furnace) may be different
depending on the intended use of it.
102
EA IsoLink IRMS System for CNSOH Operating Manual
113.
Figure
114.