Using EA IsoLink IRMS System for CNSOH
Create a Gas Configuration for a Sulfur Measurement
Prior to defining this Gas Configuration ensure that the connected IRMS has the cups for the
simultaneous detection of masses 64 and 66 and mass calibration for these cups has already been
performed.
34
For a
S measurement, a Gas Configuration must be available for the masses 64 (i.e.
32
66 (i.e.
To create a gas configuration for sulfur measurement
15. Open the Acquisition module.
16. Open theGas Configuration Editor. It is only available if no acquisition is running.
• Per default, the Gas Configuration CO
• If the Gas Configuration SO
• If the Gas Configuration SO
17. Add a new Gas Configuration.
a. Type SO
b. Select a Gas Configuration as Template, for example CO
c. Confirm by OK.
112
EA IsoLink IRMS System for CNSOH Operating Manual
16
16
S
O
O). Otherwise, it must be created as follows.
Then, proceed as follows.
for the Name.
2
already existing Gas Configurations are displayed. When creating the first Gas Configuration,
CO
is displayed.
2
is being created as the first one.
2
has already beencreated, it occurs in the list above.
2
has not been created yet, it does not occur in the list above.
2
34
16
S
. In the pull-down menu, only the
2
Thermo Scientific
16
O
O) and