Installing AI 1310/AS 1310 Autosampler
Installing the Sampler Support on the Flash IRMS Elemental Analyzer
Figure 71. Sampler Support
Guide Pivot
The support consists of a semi-circular plate resting on three spacers non adjustable in height.
The top surface of the plate is provided with two slots for the introduction of the corresponding fixing
screws. Use the guide pivot for the accommodation and the centering of the sampling unit.
Before mounting the support, you must place and fix the support bracket on the top panel of the Flash
IRMS Elemental Analyzer.
To install the sampler support on the Flash IRMS Elemental Analyzer
1. Mount the support bracket.
Note The support bracket can be installed on the left side as well as the right side of the Flash
IRMS Elemental Analyzer. Install the support bracket on the side of interest according to the
instrument configuration.
a. From the top panel of the Flash IRMS Elemental Analyzer remove the four plastic caps
b. Mount and fix the support bracket on the top panel of the Flash IRMS Elemental Analyzer by
74
EA IsoLink IRMS System for CNSOH Operating Manual
Semi-Circular Plate
covering the corresponding fixing holes. See
Figure 72. Plastic Caps Removal
using the provided fixing screws. See
Spacers
Support Bracket
Figure
72.
Figure
73.
Fixing Screws