10.5.8 Hysteresis Setting

Description
 You can set the hysteresis for alarm condition of oxygen concentration.
Use the oxygen concentration alarm 1 for the range 1, and the oxygen concentration alarm 2 for the
range 2.
Perform the setting using the percentage (%) of the range compared to the full scale.
 Settable range: 0 to 20 %
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
(6)
INZ-TN1ZKME-E
Setting the hysteresis for alarm condition of oxygen concentration to "20%" (Range 1)
Display the screen on the right in accordance with the key
operation summary and press the
Use the
key to select the hysteresis setting screen, and
press the
key.
Use the
and
key to set the hysteresis.
Press the
key to set the value.
Press the
key.
Press the
key.
The display returns to the screen on the right.
Description
key.
- 91 -
Displayed message
(LCD)
PARAMETER MENU
OXYGEN ALARM 1
OXYGEN ALARM 1
HYSTERESIS
HYSTERESIS
20 %
HYSTERESIS
20 %
OXYGEN ALARM 1
HYSTERESIS
PARAMETER MENU