Auto Calibration (Option) - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

Description
 Calibration is performed at time intervals set in advance.
 The solenoid valve is driven by contact signal to feed the standard gas for automatic calibration with
span gas and zero gas.
 "CAL" is displayed on the left of the measurement screen during automatic calibration.
 If the output signal hold is set, the output signal is held to the set value during calibration.
After the calibration, the hold is maintained until the time set in the measurement waiting time
elapses.
 For automatic calibration, it is necessary to set "10.2.2 Date and time for starting automatic
calibration (option)", "10.2.3 Cycle time setting of automatic calibration (option)", "10.2.7
Calibration gas setting".
 To perform sensor maintenance (sensor check, sensor recovery), "10.4.7 Sensor check setting for
calibration (option)" and "10.4.8 Sensor recovery setting for calibration (option)" are required.
 Refer to Sections "4.3" and "4.4" for the wiring of solenoid valves.
Sensor check
(It is executed when the function is valid.)
Sensor recovery
(It is executed when the function is valid.)
Span calibration
Zero calibration
Calibration gas valve (span)
Calibration gas valve (zero)
Analog output signal
Contact output for maintenance
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
INZ-TN1ZKME-E

7.3 Auto calibration (option)

Calibration
starting date
Setting the automatic calibration so that it is performed every four days from 13:00,
08/02/25
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key.
The auto calibration valid/invalid setting screen appears.
Use the
key to select the auto calibration valid (YES).
Press the
key to set the value.
Press the
key to set the value.
Automatic calibraiton
interval
Hold
Hold prolong
Description
key.
- 35 -
Sensor check,
recovery function
Hold
Displayed message
(LCD)
CALIBRATION MENU
SET AUTO CAL
SET AUTO CAL
AUTO CALIBRATION
AUTO CALIBRATION
YES
NO
AUTO CALIBRATION