5.3.5 Other parameters

Parameter setting
Current date and time
Contact inputs 1 to 3
Alarm contact output
Upper limit of oxygen
concentration
Range-1
Range-2
Lower limit of oxygen
concentration
Range-1
Range-2
Upper 2 limit of oxygen
concentration
Range-1
Range-2
Lower 2 limit of oxygen
concentration
Range-1
Range-2
Hysteresis
(Oxygen concentration
alarm)
Range-1
Range-2
Analog output hold
function
Maintenance hold
Error hold
Output value of analog
output hold
Maintenance hold
Error hold
Setting the value of
analog output hold
Maintenance hold
Error hold
INZ-TN1ZKME-E
Displayed message
DATE SET
00/00/01 00:00
DI 1
NONE
DO ALARM SET
ALARM NONE
HIGH ALARM
50.000 vol%
LOW ALARM
00.020 vol%
H-HIGH ALARM
55.000 vol%
L-LOW ALARM
00.010 vol%
HYSTERESIS
10 %
OUTPUT HOLD
YES
NO
OUTPUT SELECT
0%
HOLD VALUE
000 %
Range
Date and time in the
calendar
DI1 to DI3
[NONE]
[BLOW DOWN ON]
[HEATER OFF]
[PROHIBIT CAL.]
[REMOTE CAL.]
[REMOTE HOLD]
[CALCULATE REST]
[OUTPUT RANGE]
[ALARM NONE]
[HIGH ALARM]
[LOW ALARM]
[H-HIGH ALARM]
[L-LOW ALARM]
[H/L ALARM]
[HH/LL ALARM]
0.001 to 55.000 vol% in
0.001 vol% steps
0.001 to 55.000 vol% in
0.001 vol% steps
0.001 to 55.000 vol% in
0.001 vol% steps
0.001 to 55.000 vol% in
0.001 vol% steps
0 to 20 % in 1 % steps
YES or NO
[0 %] (4 mA/0V)
[100 %] (20 mA/1V)
[Last output value]
[Setting value]
0 to 100 % in 1 % steps
- 25 -
Reference
Initial value
paragraph
(00/01/01 00:00)
10.5.1
DI1
[NONE]
DI2
10.5.2
[NONE]
DI3
[NONE]
[ALARM NONE]
10.5.3
50.000 vol%
10.5.4
00.020 vol%
10.5.5
55.000 vol%
10.5.6
00.010 vol%
10.5.7
10 %
10.5.8
NO
(Analog output
10.5.9
hold function is
10.5.13
invalid.)
10.5.10
[0 %](4 mA/0V)
10.5.14
10.5.11
0 %
10.5.15