All Calibration (Option) - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

Description
 Perform sensor maintenance [sensor check (setting), sensor recovery (setting)], span and zero
calibration once for each sequentially by key operation.
 Actuate the solenoid valve attached to the exterior by the contact signal from the terminal block and
supply standard gases sequentially. Span and zero gas calibration are automatically performed.
 If the output signal hold is set, the output signal is held to the set value during calibration. After the
calibration, the hold is maintained until the time set in the measurement waiting time elapses.
 To perform sensor maintenance (sensor check, sensor recovery) during calibration, "10.4.7 Sensor
check setting for calibration (option)" and "10.4.8 Sensor recovery setting for calibration (option)"
are required.
Note that the sensor recovery is performed if it is determined to be required by the sensor check.
 Refer to Sections "4.3" and "4.4" for the wiring of solenoid valves.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
How to cancle
 Press the
key to cancel the operation.
INZ-TN1ZKME-E

7.5 All calibration (option)

Executes all calibration.
Display the screen on the right in accordance with the key
operation summary and press the
performing screen appears.
Press the
key to perform all calibration.
The value of the concentration of oxygen and the cell
electromotive force are displayed while executing the all
calibration.
After the all calibration is completed, the display returns to the
screen on the right.
Description
key, the all calibration
- 39 -
Displayed message
(LCD)
CALIBRATION MENU
ALL CALIBRATION
ALL CALIBRATION
START
ALL CALIBRATION
20.61 % 000.1 mV
CALIBRATION MENU
ALL CALIBRATION