Performing All Calibration (Option) - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

10.2.4 Performing all calibration (option)

Description
 You can perform all calibration on the screen using this function.
Zero calibration is automatically performed after the span calibration.
 If the execution of the treatment is set in "10.4.7 Sensor check setting for calibration (option)" and
"10.4.8 Sensor recovery setting for calibration (option)," sensor check and sensor recovery are
performed before the calibration.
Note that the sensor recovery is performed if it is determined to be required at the sensor check.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
How to cancel
 Press the
key to cancel the operation.
Performing all calibration on the screen
Description
Display the screen on the right in accordance with the key
operation summary and press the
calibration performing screen appears.
Press the
key to perform all calibration.
Oxygen concentration value and cell electromotive force are
displayed during all calibration.
After the all calibration is completed, the display returns to
the screen on the right.
key, the all
- 56 -
Displayed message
(LCD)
CALIBRATION MENU
ALL CALIBRATION
ALL CALIBRATION
START
ALL CALIBRATION
20.61 % 000.1 mV
CALIBRATION MENU
ALL CALIBRATION
INZ-TN1ZKME-E

Advertisement

Table of Contents
loading

Table of Contents