10.5.14 Hold value setting (error hold)

Description
 Using this function, you can set (select) the output value of analog output signal when the analog
output hold function (error hold) is enabled.
 If the maintenance mode is set to "Valid," analog output signal is held at the value set in this
procedure.
 Settable range: Select one of the following.
(1) [0%]
(2) [100%]
(3) [Last value]
(4) [Setting value] : Held at the value set as the "10.5.15 Setting of hold setting
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
: Held at 0% (4 mA/0 V).
: Held at 100% (20 mA/1 V).
: Held at the value immediately before the value for analog hold.
value (error hold)".
Setting the output value of analog output hold to "0%"
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key to display the screen on the right and press
the
key.
The analog output hold value setting screen appears.
Use the
key to select the hold value.
Press the
key to set the value.
Press the
key.
Press the
key to return to the screen on the right.