Date And Time For Starting Automatic Calibration (Option) - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

10.2.2 Date and time for starting automatic calibration (option)

Description
 You can set the date and time for starting automatic calibration using this function.
Automatic calibration is performed in a specified cycle from a specified date and time.
 If it is invalid, the automatic calibration does not start at a specified date and time.
 Settable range: date and time in the future in the calendar
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
Note
 You cannot change the setting value during automatic calibration or remote calibration.
 Check that "Current date and time setting" in the parameter menu is properly set.
Setting the automatic calibration so that it is performed from 13:00, 08/02/25
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key to display the screen on the right and press
the
key.
The auto calibration starting date and time setting screen
appears.
Use the
and
key to set the auto calibration starting
date and time.
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
Description
key.
- 54 -
Displayed message
(LCD)
CALIBRATION MENU
SET AUTO CAL
SET AUTO CAL
START DATE
START DATE
08/01/01 00:00
START DATE
08/02/25 13:00
SET AUTO CAL
START DATE
INZ-TN1ZKME-E