Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

2.1.2 Terminal block

O
O
sensor
thermocouple
2
input
Exclusive
cable
1
2
+
O
2
Ground
+
TC2
terminal
M4
15
Thermocouple
input type R
(not supplied)
COMMUNICATION TERMINAL (TM2) /INSERTION TERMINAL
INZ-TN1ZKME-E
EXTERNAL TERMINAL (TM1) /M3
sensor
2
input
Analog
Fault
output
contact output
4-20mA DC
250V AC/3A
or 0-1V DC
30V DC/3A
3
4
5
6
7
AO
+
+
TC1
or
FAULT
ALARM
DI1
DICOM
DI2
DI3
16
18
19
17
20
Contact input
Short circuit : on (Current 5mA)
Open
: off (Voltage 5V)
Terminal number
1
2
RS232C
TXD
RXD
RS485
TRX+
TRX–
O
2
Blow contact
heater power
output
output
250V AC/3A
COM
30V DC/3A
8
9
10
11
12
ZV
BLOW
ZERO
GAS
HEATER
MAINTE
COM
21
22
23
24
25
Alarm
Maintenance
Calibrating gas
contact output
contact output
contact output
250V AC/3A
250V AC/3A
250V AC/3A
30V DC/3A
30V DC/3A
30V DC/3A
Remarks
3
GND
Standard
GND
Option
- 3 -
sensor
13
14
SPAN
GAS
AC
N
SV
L
26
27
28
AC
Power supply
input
100 to120V AC
200 to 240V AC
Note 1) The heater power supply is the same
as the converter power supply.
Note 2) Be sure to connect the shield of the 
cable to the ground in the main body.