10.3.3 Automatic blowdown cycle setting

Description
 You can set the automatic blowdown cycle using this function.
The cycle starts from a specified date and time for automatic blowdown.
 Settable range: 00h 00m to 99h 59m (m: 00 to 59)
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
Note
 You cannot change the setting value during automatic blowdown or remote blowdown.
 Set the blowdown cycle value larger than the blowdown time.
INZ-TN1ZKME-E
Setting automatic blowdown cycle to 24 hours.
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key to display the screen on the right and press
the
key.
The auto blowdown cycle setting screen appears.
Use the
and
key to select the auto blowdown cycle.
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
Description
key.
- 65 -
Displayed message
(LCD)
BLOW DOWN MENU
SET AUTO BLOW
SET AUTO BLOW
AUTO BLOW CYCLE
AUTO BLOW CYCLE
24 h 00 m
AUTO BLOW CYCLE
24 h 00 m
SET AUTO BLOW
AUTO BLOW CYCLE