Contact Input Setting - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

10.5.2 Contact input setting

Description
 You can set the functions for the contact inputs 1 to 3 using this function.
 Settable range: Select one of the following
(1) [NONE]
(2) [BLOW DOWN ON] : Performs blowdown by contact input.
(3) [HEATER OFF]
(4) [PROHIBIT CAL.]
(5) [REMOTE CAL.]
(6) [REMOTE HOLD]
(7) [CALCULATE REST] : Resets maximum and minimum calculations of O
(8) [OUTPUT RANGE] : Switches the range by contact input.
Note) The functions other than "NONE" cannot be set for multiple contacts.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
(6)
INZ-TN1ZKME-E
Setting the blowdown function for the contact input 1
Display the screen on the right in accordance with the key
operation summary and press the
Use the
key to select the contact input 1 setting screen.
Press the
key to set the value.
(Also follow this procedure for the contact inputs 2 and 3.)
Use the
key to select the function for contact input 1.
The item selected is highlighted.
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
: Performs no treatment by contact input.
(Switch OFF to ON to perform blowdown.)
: Turn off the heater by contact input.
(OFF/ON:Heater ON/Heater OFF)
: Sets if calibration is prohibited or valid by contact
input.
(OFF/ON: Calibration is valid/prohibited.)
: Performs all calibration by contact input.
(Switch OFF to ON to perform calibration.)
: Holds the AO by contact input.
(OFF/ON: not held/held)
contact input.
(Switch OFF to ON to perform calibration.)
(OFF/ON: Range-1/Range-2)
Description
key.
- 85 -
by
2
Displayed message
(LCD)
PARAMETER MENU
DIGITAL INPUT
DIGITAL INPUT
DI 1
DI 1
NONE
DI 1
BLOW DOWN ON
DI 1
BLOW DOWN ON
DIGITAL INPUT
DI 1