10.5.15 Setting of hold setting value (error hold)

Description
 Using this function, you can set the output value of analog output signal to an arbitrary value when the
analog output hold function (error hold) is enabled.
 This function is enabled if "Setting value" is set at "10.5.14 Hold value setting (error hold)."
 Set the output value of analog output signal as a percentage (%) of the full-scale value of the display
range. 0% is equivalent to 0 vol% (4 mA/0 V) and 100 % is to the full-scale value (20 mA/1 V).
 Settable range: 0 to 100 %
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
INZ-TN1ZKME-E
Setting the output value of analog output hold to "000%"
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key to display the screen on the right and press
the
key.
The hold setting value setting screen appears.
Use the
and
key to set the hold value.
Press the
key to set the value.
Press the
key.
Press the
key to return to the screen on the right.
Description
key.
- 99 -
Displayed message
(LCD)
PARAMETER MENU
AO HOLD (ERROR)
AO HOLD (ERROR)
HOLD VALUE
HOLD VALUE
000 %
HOLD VALUE
000 %
AO HOLD (ERROR)
HOLD VALUE