Setting Of Measurement Recovery Time (Maintenance Hold) - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

10.5.12 Setting of measurement recovery time (maintenance
hold)
Description
 This function is to set the recovery time (hold of extension) from the analog output hold function
(maintenance hold).
 Settable range: 0 to 300 sec.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
Setting the time for extension of hold to "10 seconds"
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key to display the screen on the right and press
the
key.
The measurement recovery time setting screen appears.
Use the
and
key to set the measurement recovery
time.
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
Description
key.
- 96 -
Displayed message
(LCD)
PARAMETER MENU
AO HOLD (MAINTE)
AO HOLD (MAINTE)
MEAS. WAIT TIME
MEAS. WAIT TIME
010 S
MEAS. WAIT TIME
010 S
AO HOLD (MAINTE)
MEAS. WAIT TIME
INZ-TN1ZKME-E