6. SETTING AND CALIBRATION

6.1 Calibration setting

Calibration setting is made to select the concentration at calibration and calibration operation for the
range.
6.1.1 Setting of calibration concentration
pressing the
press the
ENT
(2) In the calibration setting item screen that
appears, move the cursor to "Calibration
and then press the
make the setting by pressing the
or the
key.
ENT
(4) Enter zero and span calibration gas
concentration by using the
key, and then press the
To close "
To close "Calibration setting" or terminate the
operation, press the
ESC
the previous screen.
22
or the
key and then
key.
or the
key.
ENT
,
or the
key.
ENT
"
User Mode
Setting about Calibration
Alarm Setting
Setting of Auto Calibration
Changeover of Range
Parameter Setting
Cal. Settings
key,
Calibration Value
About Calibration Range
Cal. Settings
Cal. Value
H
2
RANGE
0–3vol%
0–10vol%
Select an item
with UP/DOWN and ENT
Back with ESC
Select an item
with UP/DOWN and ENT
Back with ESC
Select setting value
ZERO
SPAN
0 0 . 0 0
2 . 9 1 0
0 0 . 0 0
0 9 . 8 1