Decimal Point Position Setting Screen - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

10.1.2 Decimal point position setting screen

Description
 You can set the decimal point position of full scale for oxygen concentration display using this
function.
 Settable range: Select one of the following.
(1) "00.00": Displayed with two-digit integer and two decimal places.
(2) "0.000": Displayed with one-digit integer and three decimal places.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
Note
 If changing "0.000" to "00.00," "25.00" is set as the full scale value.
 If changing "00.00" to "0.000," "5.000" is set as the full scale value.
Setting the display of two-digit integer and two decimal places (Range 1)
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key.
The decimal point position setting screen appears.
Use the
key to select the two-digit integer and two
decimal places.
Press the
key to set the value.
Press the
key.
When it is fixed, the display returns to the screen on the right.
Description
key.
- 50 -
Displayed message
(LCD)
MEASURE MENU
OUTPUT RANGE
RANGE 1
DECIMAL POINT
DECIMAL POINT
00.00
DECIMAL POINT
00.00
RANGE 1
DECIMAL POINT
INZ-TN1ZKME-E