10.2.8 Calibration waiting setting (option)

Description
 Set the waiting time from supply of calibration gas to start of calibration.
(Set the time so that the calibration gas becomes stable before the calibration.)
 Settable range: 10 to 300sec.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
Note
 You cannot change the setting value during automatic calibration or remote calibration.
Setting the wait time to start of calibration to 20 seconds
Description
Display the screen on the right in accordance with the key
operation summary and press the
The currently set content is displayed.
Use the
and
key to change the wait time.
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
key.
- 60 -
Displayed message
(LCD)
CALIBRATION MENU
CAL. WAIT TIME
CAL. WAIT TIME
020 S
CAL. WAIT TIME
020 S
CALIBRATION MENU
CAL. WAIT TIME
INZ-TN1ZKME-E