10.4.12 Maintenance mode setting

Description
 You can set the maintenance mode to valid or invalid with this function.
 If the maintenance mode is set to valid, the analog output signal is held at the set value (see "10.5.10
Hold value setting.") and the contact output for maintenance of the external contact is on. The data
portion of the measurement screen flickers.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
Note
 If an error occurs while the maintenance mode is enabled, error handling is prioritized.
 If the analog output hold function (error hold) is enabled, the analog output signal is held at the value
set at the hold value setting (error hold).
 The data portion of the measurement screen flickers and is highlighted.
INZ-TN1ZKME-E
Setting the maintenance mode to valid
Display the screen on the right in accordance with the key
operation summary and press the
mode setting screen appears.
Use the
key to select the maintenance mode valid (YES).
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
Description
key , the maintenance
- 81 -
Displayed message
(LCD)
MAINTE MENU
MAINTENANCE MODE
MAINTENANCE MODE
YES
NO
MAINTENANCE MODE
YES
MAINTE MENU
MAINTENANCE MODE