Installation; Installation Site - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

DANGER
 Although this unit is the explosion-proof type, do not use it in the place (zone 0) where explosive
gases are always exist. Otherwise serious accidents such as explosion or fire may result.
CAUTION
 For installation, observe the rule on it given in the instruction manual and select a place where the
weight of converter can be endured. Installation at an unsuited place may cause turnover or fall and
there is a risk of injury.
 Before transport, fix the door so that it will not open. Otherwise, the casing may be separated and fall
to cause an injury.
 During installation work, care should be taken to keep the unit free from entry of cable chips or other
foreign objects. Otherwise, it may cause fire, failure or malfunction.
Install the converter in a place that satisfies the following conditions.
(1) Space for periodic inspection and wiring work is available.
(2) Vibration, dust, dirt, and humidity are minimal.
(3) A place not subjected to radiated heat from a heating furnace, etc.
(4) The atmosphere is non-corrosive.
(5) Away from electrical devices that may cause noise trouble (such as motor and transformer), and
equipment that may cause electromagnetic or electrostatic induction trouble.
(6) A place where ambient temperature and humidity are -20 to +55°C and 95%RH or less.
Secure at least 100 mm of space between the converter and nearby wall. Also secure a space of opening the
front cover for maintenance.
Secure a cable wiring space under the case.
INZ-TN1ZKME-E

3. INSTALLATION

3.1 Installation site

100
or more
526 or more
- 7 -
100
326
or more
Unit : mm