10.4.8 Sensor recovery setting for calibration (option)

Description
 You can set if a sensor recovery is performed for calibration using this function.
Note that the sensor recovery is performed only if it is determined to be required at the sensor check.
 This function is performed only if valid is selected in the sensor check setting for calibration.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
(6)
INZ-TN1ZKME-E
Making a setting so that sensor recovery is performed for calibration
Display the screen on the right in accordance with the key
operation summary and press the
Use the
key to select the sensor recovery setting screen
for calibration.
Use the
key to select the sensor recovery valid (YES).
Press the
key to set the value.
After the setting is completed, the display returns to the screen
on the right.
Press the
key again to return to the screen on the right.
Description
key.
- 77 -
Displayed message
(LCD)
MAINTE MENU
CAL. CELL MAINTE
CAL. CELL MAINTE
SENSOR RECOVER
SENSOR RECOVER
YES
NO
SENSOR RECOVER
YES
CAL. CELL MAINTE
SENSOR RECOVER
MAINTE MENU
CAL. CELL MAINTE