10.2.3 Cycle time setting of automatic calibration (option)

Description
 You can set the automatic calibration cycle using this function.
The cycle starts from a specified date and time for automatic calibration.
 Settable range: 00d 00h to 99d 23h (h: 00 to 23)
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
Note
 You cannot change the setting value during automatic calibration or remote calibration.
INZ-TN1ZKME-E
Setting the automatic calibration so that it is performed every four days
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key to display the screen on the right and press
the
key.
The cycle time setting of automatic calibration appears.
Use the
and
key to set the auto calibration starting
date and time.
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
Description
key.
- 55 -
Displayed message
(LCD)
CALIBRATION MENU
SET AUTO CAL
SET AUTO CAL
AUTO CAL. CYCLE
AUTO CAL. CYCLE
04 d 00 h
AUTO CAL. CYCLE
04 d 00 h
SET AUTO CAL
AUTO CAL. CYCLE