Initial Parameter Value Table; Parameters Related To Measurement; Parameters Related To Calibration - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

5.3.1 Parameters related to measurement

Parameter setting
Display range
Decimal point position
Range1
Range2
Full scale
Range1
Range2
Calculation time of
maximum and minimum
values

5.3.2 Parameters related to calibration

Parameter setting
Auto calibration function
(Displayed if the option is
provided.)
Date and time for starting
automatic calibration
(Displayed if the option is
provided.)
Automatic calibration
cycle time
(Displayed if the option is
provided.)
Calibration gas
concentration-1
calibration gas
concentration-2
Calibration wait time
Calibration range setting
INZ-TN1ZKME-E

5.3 Initial parameter value table

Displayed message
OUTPUT RANGE
RANGE1
RANGE2
DECIMAL POINT
00.00
FULL SCALE
25.00
CALCULATE TIME
024 h
Displayed message
AUTO CALIBRATION
YES
NO
START DATE
99/01/01 00:00
AUTO CAL. CYCLE
07d 00h
SPAN
ZERO
20.600% 02.000%
CAL. WAIT TIME
020 s
ABBOUT CAL. RANGE
BOTH
CURRENT
Range
Range1 or Range2
[00.00]
[0.000]
2 to 50 in 1 vol% steps
0 to 240 hour in 1-hour
steps
Range
YES or NO
Date and time in the
future in the calendar
00d 00h to 99d23h
(h: 00 to 23)
Span:
00.010 to 50.000 vol%
Zero:
00.010 to 25.000 vol%
in 0.001 vol% steps
10 to 300 sec.
in 1 sec. steps
Set calibration range
Current or both range
- 23 -
Reference
Initial value
paragraph
Range-1
10.1.1
[00.00]
10.1.2
25.00 vol%
10.1.3
24 hour
10.1.4
Reference
Initial value
paragraph
Invalid
(Auto calibration
10.2.1
function: Invalid)
99/01/01 00:00
10.2.2
07d 00h
10.2.3
Span:
20.600 vol%
10.2.7
Zero:
02.000 vol%
20 sec.
10.2.8
BOTH
10.2.10