Calibration Gas Setting - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

10.2.7 Calibration gas setting

Description
 Set calibration gas concentration (span/zero calibration gas concentrations).
Use the calibration gas concentration 1 for the range 1, and the calibration gas concentration 2 for the
range 2.
 Use normal air (atmosphere) as a span calibration gas and set its concentration to 20.600% O
 Settable range: Span calibration gas 00.010 to 50.000 %O
Zero calibration gas 00.010 to 25.000 %O
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
Note
 You cannot change the setting value during automatic calibration or remote calibration.
 Set with span calibration gas concentrations ≥ zero calibration gas concentrations.
INZ-TN1ZKME-E
Setting the span/zero calibration gas concentrations (Range 1)
Display the screen on the right in accordance with the key
operation summary and press the
The set content is displayed now.
Use the
and
key to change the calibration gas
concentrations.
Press the
key to set the value.
The set content is displayed.
Press the
key.
The display returns to the screen on the right.
/N
2
2
/N
2
2
Description
key.
- 59 -
/N
.
2
2
Displayed message
(LCD)
CALIBRATION MENU
CAL. GAS (RANGE 1)
SPAN
ZERO
20.600% 02.000%
SPAN
ZERO
20.600% 02.000%
SPAN
ZERO
20.600% 02.000%
CALIBRATION MENU
CAL. GAS (RANGE 1)