10.5.20 Fuel coefficient setting (option)

Description
 You can set the fuel coefficient [k], which is for calculation of combustion efficiency, using this
function.
 Settable range : 0.00 to 1.99
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
Calculation and display of combustion efficiency
 Using the following formula, calculate the combustion efficiency E.
Combustion efficiency E [ % ] =100 – Fuel coefficient [k]
Standard oxygen concentration value [vol%]-Measured oxygen concentration value [vol%]
Note 1: Fuel coefficient [k] is according to the fuel used.
Refer to the table below.
Note 2: Temperature of flue gases should be measured separately.
Connect the electromotive force signal of the thermocouple type temperature detector (R-
type) to this unit.
Note 3: The standard oxygen concentration value is to be 20.8 [vol % ].
Note 4: When the internal calculation result of the combustion efficiency [E]
exceeds 0 [%] ≤ E ≤ 100 [%], % is displayed.
Setting the fuel coefficient [k] to 0.73.
Display the screen on the right in accordance with the key
operation summary and press the
Use the
key and the
Press the
key to fix.
Press the
key to return to the screen on the right.
Temperature of flue gases [K]
Description
key.
key to input the fuel coefficient.
Natural gas
Oil
Bituminous coal
- 104 -
Displayed message
(LCD)
PARAMETER MENU
FUEL COEFF.
FUEL COEFF.
0.73
FUEL COEFF.
0.73
PARAMETER MENU
FUEL COEFF.
k = 0.66
k = 0.7
k = 0.73
INZ-TN1ZKME-E