CAUTION ON SAFETY To operate the analyzer properly, be sure to read “Caution on Safety” carefully. DANGER CAUTION damage. PROHIBITION CAUTION Caution on installation and transport of gas analyzer DANGER CAUTION INZ-TN2ZPB-E...
Page 4
Caution on piping Be sure to observe the following precautions while installing DANGER piping. Improper piping may result in gas leakage. If the leaking gas contains a toxic component, serious acci- dents may result. If it contains combustible gases, explosion Caution on wiring CAUTION Caution on use...
Page 5
Caution on use PROHIBITION Caution on maintenance and check DANGER Be sure to observe the following to perform work safely, CAUTION avoiding electric shock or injury. Others CAUTION INZ-TN2ZPB-E...
WARRANTY AND MAINTENANCE Scope of application Operating conditions and environment Precautions and prohibitions Warranty 4-1. Period of warranty 4-2. Scope of warranty from: INZ-TN2ZPB-E...
Failure diagnosis Service life Maintenance plan Maintenance Preventive Daily inspection maintenance Periodic inspection Corrective Troubleshooting maintenance INZ-TN2ZPB-E...
Page 8
Limited-life parts and consumable parts Spare parts and accessories accessories. 10. Period for repair and provision of spare parts after product dis- continuation (maintenance period) INZ-TN2ZPB-E...
2. NAME OF DELIVERED ITEMS AND EACH PARTS Analyzer: 1 unit Standard: IEC127-2 Size: ø5 × 20mm Fuse: 2 pcs Rating: 250V/2A delay type Part No.: R75796N17 25 pin D-sub connector (male) Analog output connector: 1 Part No.: R77256N262 Fixing screws: 2 M2.6 ×...
2.2 Name and description of analyzer (3) Flowmeter (unit mm) < Front panel > (2) Display/operation panel (1) Power switch (12) Analog output connector (A / O) (D-sub25 pin) (4) Purge gas inlet (11) Communication connector (RS-485) (5) Sampling gas inlet (6) Sampling gas and (10) External input connector reference gas outlet...
3.2 Installation 3.2.1 Installation of analyzer main frame (Unit : mm) Type External dimensions Mounting dimensions Mounting method 19-inch rack mounting 450 or more Support “A” : 57.2 (EIA) INZ-TN2ZPB-E...
Page 16
3.3 Piping CAUTION coupling. Purge gas inlet Sampling gas inlet 1 Reference gas inlet Sampling gas and reference gas outlet 1 Sampling gas inlet 2 For special case only Sampling gas outlet 2 INZ-TN2ZPB-E...
Page 17
Internal piping diagram Optical unit 2 (IR Unit 2) Optical unit 1 (IR Unit 1) (Left) (Right) Built-in sensor Sampling cell REFERENCE INLET1 OUTLET1 Note) 2 cells may be used Sampling gas inlet Sample gas outlet/ by combination of Reference gas outlet range.
3.4 Sampling 3.4.1 Conditions of sampe gas consider- 3.4.3 Preparation of standard gas mea- sensor sensor Zero gas Dry air mea- 3.4.4 Purging of instrument inside INZ-TN2ZPB-E...
3.4.5 Pressure at sampling gas outlet (1) Gas extractor ZPB unit (7) Membrane filter (8) Flowmeter ø10/ø8 (4) Gas aspirator Teflon Air filter tube analyzer (11) NO /NO converter (Wet N2 or main frame (for NO measurement) 3-way atmospheric air)
Page 20
3.4.7 Gas requirements for measurement and calibration From reference From sample gas Remarks gas inlet inlet Zero calibration Dry N or dry air Dry N or dry air Use dry gas (cylinder) for Calibration both zero and span cali- Span calibration Dry N or dry air Dry span gas...
Page 21
3.5 Wiring CAUTION CAUTION External input connector (A/I) Communication connector (RS-485) Analog output connector (A/O) Power terminal block Digital input/output connector Power inlet (DIO 1 to 3) INZ-TN2ZPB-E...
(1) Power supply <Terminal block for power supply> <Power inlet> 3 2 1 Attached ferrite core Power inlet AC100 to 240V Earth cable Grounding 2-pole plug CAUTION When noise source is in the vicinity Main unit power supply Varistor or spark quencher Install (connect) near the source.
Page 23
(2) Analog output signal: Analog output connector (A/O) < Analog output > A/O connector AO1+ AO1- AO2+ AO2- AO3+ D-sub 25-pin female AO3- AO4+ Note) Display Ch number is same AO4- as the AO number under AO5+ AO5- standard specifications. AO6+ AO6- AO7+...
Page 24
(3) O sensor input: External input connector (A/I) or more) < External input > A/I connector (O sensor input) ordered. full scale analyzer is Turn To O sensor − Solder the signal cable to the terminal. INZ-TN2ZPB-E...
Page 25
(4) Contact input/output (DIO): digital input/output connector (DIO1 to 3) < Digital input/output > Connector for DIO 1 to 3 (option) DIO1 DIO2 DIO3 connector connector connector DI1+ DI4+ DI7+ DI1- DI4- DI7- Digital input DI2+ DI5+ DI8+ OFF : 0V DI2- DI5- DI8-...
Page 27
2) In case of automatic calibration (example shown in Section 6.4.1, Auto calibration) Automatic Ch1 span Ch3 span calibration Ch5 span calibration calibration start calibration Ch4 span Ch2 span Zero calibration calibration calibration Zero gas Zero calibration output 350 s Ch1 span calibration output Ch1 span gas 350 s...
4. OPERATION 4.1 Preparation for operation (1) Piping and wiring check 4.2 Warm-up operation and regular operation (1) Operation procedure CAUTION When in warm-up, the concentration reading may be beyond the upper limit of range. But, it is not an error. INZ-TN2ZPB-E...
5. DESCRIPTION OF DISPLAY AND OPERATION PANELS 5.1 Name and description of operation panel Display unit Operation panel (7) ZERO key (8) SPAN key MODE ZERO SPAN (1) MODE key (4) DOWN key (2) SIDE key (3) UP key (5) ESC key Name Description Name...
5.3 Outline of display screen (1) Measurement mode screen (appears when the power is turned ON) 0-50 0-50 0-50 0-50 0.0 0 Scroll vol% 0-20 0-50 0-50 0-50 0.0 0 vol% vol% 0-25 vol% 0-25 INZ-TN2ZPB-E...
Page 32
Instantaneous value and concentration value: corrected concentration values: × corrected concentration average value: aver- (2) Setting/selection screen Message display area Status display area display area Cursor INZ-TN2ZPB-E...
6. SETTING AND CALIBRATION 6.1 Switch of range Measurement Mode Screen 6.1.1 Setting of range switch mode MODE MODE User Mode Select an item with UP/DOWN and ENT Back with ESC Switch Ranges Calibration Parameters key. Alarm Setting Setting of Auto Calibration Setting of Auto Zero Calibration Peak Alarm Setting Parameter Setting...
6.2 Calibration setting Cal. Parameters Select an item with UP/DOWN and ENT Back with ESC Calibration Value 6.2.1 Setting of calibration concentration About ZERO Calibration About Calibration Range About Calibration Components / Range Cal. Settings Select setting value Cal. Value RANGE ZERO SPAN...
Page 38
Cursor for setting value value increases or decreases. By pressing Cal. Settings Set calibration value Cal. Value RANGE ZERO SPAN 0–50.00ppm +0000.0 50.00 0–500.0ppm +00000 500.0 0–50.00ppm +0000.0 50.00 Note) Enter settings that correspond to 0–500.0ppm +00000 500.0 each range. If zirconia type is used 0–10.00ppm +000.00 10.00...
6.2.2 Setting of manual zero calibration Cell. Settings Set each or at once Ch ZERO Call. at ZERO Calibration Range1 0–50.00 at once Range2 0–500.0 Range1 0–50.00 at once Range2 0–500.0 Range1 0–10.00 vol% at once Range2 0–20.00 vol% Range1 0–50.00 at once Range2...
Page 40
Manual Calibration screen When setting all components to “each”: When setting all components to “at once”: ZERO Call. ENT : Go on Calibration ZERO Call. ENT : Go on Calibration of selected Ch of selected Ch ESC : Not calibration ESC : Not calibration -2.1 Range1...
6.2.3 Setting of calibration range Cell. Settings Set calibration range Cell. Range current or both range Range1 0–50.00 both Range2 0–500.0 Range1 0–50.00 current Range2 0–500.0 Range1 0–10.00 vol% current Range2 0–20.00 vol% Range1 0–50.00 both Range2 0–500.0 key. Range1 0–10.00 vol% current...
6.2.4 Setting of auto calibration component/range Cell. Settings Select a range for Auto Cal. auto calibration Range1 0–50.00 enable Range2 0–500.0 Range1 0–50.00 enable Range2 0–500.0 Range1 0–10.00 vol% enable Range2 0–20.00 vol% Range1 0–50.00 enable Range2 0–500.0 Range1 0–10.00 vol% enable Range2...
Page 43
To close the setting Press the key to exit automatic calibration component/range setting, and the previous screen appears. Operation by setting Auto calibration is performed under the following rules. 1. Zero calibration is performed at the same time, for the Ch (component) in which “enable” is selected at the time of auto calibration and auto zero calibration.
6.3 Alarm setting 6.3.1 Setting of alarm values Alarm Setting Select Alarm No. or Hysteresis setting Alarm-1 key. Press Alarm-2 key. Alarm-3 Alarm-4 Alarm-5 Hysteresis Alarm Setting Select an item with UP/DOWN and ENT Alarm-1 Back with ESC key. Channel CAUTION H-Limit Range 1 50.00...
Page 45
Description of setting items The alarm contact assigned the same number as the alarm is operated accordingly. Channel: Channel setting targeted for issuance of alarm. One Ch No. can be selected for multiple alarms. H-Limit value: Sets the high limit value (concentration) of alarm. L-Limit value: Sets the low limit value (concentration) of alarm.
6.3.2 Hysteresis setting esis. Alarm Setting Set Hysteresis 0 to 20%FS available Alarm-1 Alarm-2 Alarm-3 Alarm-4 Alarm-5 Hysteresis End of Hysteresis Setting To close "Hysteresis Setting" To close the “Hysteresis Setting” or cancel the mode midway, press the key. A previous screen will return. CAUTION Setting range The hysteresis is common to all alarms...
6.3.3 Peak alarm setting Measurement Mode Screen MODE MODE User Mode Select an item with UP/DOWN and ENT key. Back with ESC key. Switch Ranges Calibration Parameters Alarm Setting Setting of Auto Calibration key. Setting of Auto Zero Calibration Peak Alarm Setting Parameter Setting Peak Alarm Select setting item...
Page 48
Setting range Alarm value : 10 to 1000 ppm 5 ppm step (initial value: 500 ppm) Alarm count : 1 to 99 times (per hour) (initial value: 5 times) Hysteresis : 0 to 20 % of full scale (initial value: 0% of full scale) [% full scale] represents the percentage with the CO range regarded as 100%.
6.4 Setting of auto calibration 6.4.1 Auto calibration Set Auto Cal. Select setting item Start Time 12:00 Cycle key. Flow Time ON / OFF Time : MON 12:34 Auto Calibration Run Set Auto Cal. Set Start Time Start Time 12:00 Press the Cycle key, and...
Page 50
Set Auto Cal. Set flow time of calibration gas 60 to 900 sec ZERO 350 sec. Ch1 Span 350 sec. key. Ch2 Span 350 sec. Ch3 Span 350 sec. Ch4 Span 300 sec. Ch5 Span 300 sec. Ex. time 300 sec. key.
cases. Example 12:00 Start Time Cycle 350 sec Flow Time Zero 350 sec Ch1 Span 350 sec Ch2 Span 350 sec Ch3 Span 300 sec Ch4 Span 300 sec Ch5 Span 300 sec EX. time ON/OFF In case where auto calibration is carried out at the above setting. Sunday Monday Tuesday...
6.4.2 Forced run/stop of auto calibration 6.4.2.1 Execution of auto calibration (only once) Set Auto Cal. Auto Cal. Run ENT : Run / Stop ESC : Cancel key. Start Time 12:00 Cycle Flow Time ON / OFF Time : MON 12:34 Auto Calibration 6.4.2.2 Forced stop of auto calibration Set Auto Cal.
6.5 Setting of auto zero calibration 6.5.1 Auto zero calibration Set Auto Select setting item Zero Cal. Start Time 12:00 Cycle key. Flow Time sec. ON / OFF Time : MON 12:34 Auto Zero Calibration Run Set Auto Set Start Time Zero Cal.
Page 55
Example Start time 12:00 Cycle hour Flow time ON/OFF In case where auto zero calibration is carried out at the above setting. Sunday Monday Monday 12:00 0:00 12:00 Cycle : Auto zero calibration Replace- Zero ment calibration time 300sec 300sec Gas replacement time after calibration is the same as the flow time.
6.5.2 Forced run/stop of auto zero calibration 6.5.2.1 Execution of auto zero calibration (only once) Set Auto Auto zero Run ENT : Run / Stop Zero Cal. ESC : Cansel key. key. Start Time 12:00 Cycle Flow Time sec. ON / OFF Time : MON 12:34 cancel.
Page 57
“Auto Zero Calibration” screen Example 0 3 . 0 0 0 0 0 . 2 1 0 2 CAUTION During auto zero calibration, any key operation is not permitted other than operations such as key lock ON/OFF and “Forced stop of auto zero calibration”. When the key lock is set at ON, even the “Forced stop of auto zero calibration”...
6.6 Parameter setting Description of setting items Parameter Select setting item Current Time 12/01/11 WED 13:50 key. Key Lock Output Hold Current Response Time Average Period Backlight Timer 05 min Contrast To Maintenance Mode 0000 Parameter Set day of week Current Time 12/01/11 13:50...
Page 59
Setting Range Output Hold Manual calibration ZERO Press the key to flow gas. SPAN To execute calibration, press the key. Calibration Hold extending time Output hold Time set to gas flow time (See Section 6.4) b. Auto calibration Auto calibration start Span calibration end Calibration Hold extending time.
Page 60
Parameter Select Hold ON or OFF Current Time 12/01/11 WED 13:50 Key Lock Output Hold Current Response Time Average Period Backlight Timer 05 min Contrast To Maintenance Mode 0000 Parameter Select Hold setting key. Current Time 12/01/11 WED 13:50 Key Lock Output Hold Setting Response Time...
Parameter Set Hold value 0 to 100%FS Hold key. key. Meaning of setting The setting is expressed as 1/1 full scale range for both respective ranges. When 0 to 1000 ppm is selected as the range, and 10% FS is selected as hold setting, the output equivalent to 100 ppm is End of Hold Setting held irrespective of the measure-ment value at...
Page 62
Average period Parameter Select Ch No. average. Average Period hour Ch10 hour Ch11 hour average value. (Press- Ch12 hour Reset AV. Output Reset Average value reset Contact input flow With input (hold at least 1.5 sec.) Without input Reset input So long as with input, resetting lasts.
Backlight Timer Parameter Set Backlight Timer ON or OFF Current Time 12/01/11 WED 13:50 Key Lock Output Hold Setting off. Response Time Average Period Backlight Timer 05 min Contrast To Maintenance Mode 0000 Contrast Parameter key. key. Current Time 12/01/11 WED 13:50 Key Lock Output Hold Setting...
6.7 Maintenance mode Maintenance Select operating item Mode 1. Sensor Input Value 2. Error Log 3. Cal. Log 4. Output Adj. 5. Other Parameter key. 6. To Factory Mode played. Note) “To Factory Mode” is used for our service engineers only. Each “Maintenance”...
Page 65
Description of Calibration Log screen Maintenance Select Ch No. Cal. Log Past calibration history is displayed. Sensor input value, concentration value, and the date when zero/span calibration is performed are logged. The 10 newest calibration data are logged by each component. Move the cursor to Clear Calibration Log and press the key, and the calibration log is...
Page 66
Description of output adjustment screen Maintenance Adjust OUTPUT Analog output adjustment screen. ZERO and SPAN Mode Connect the digital multi meter to the output Output Adj. terminal corresponding to the number of OUT Zero Span Zero Span to be adjusted, and adjust the value so that 4mA 00600 03700 00600 03700 or 0V is output at zero and 20mA or 1V is...
Page 67
Description of each setting screen Maintenance Set Password Mode Password Set : Set the password used to move Setting from the parameter setting screen to the maintenance mode. Password Set 2465 Arbitrary 4-digit number can be ref. Value 12% O limit 20% O selected.
Page 68
Maintenance Select an item Mode setting Password set 2465 ref. Value 12% O limit 20% O Station No. 01 Range setting key. Maintenance Select Ch No. Mode Range set key. Maintenance Select range or Mode range num. Range set key. Ch1 NOx MIN range 50.00...
Page 69
6.7.1 Sample switch setting Maintenance Select operating item Mode Sample Switch Gas Flow Time Interference Coefficient (meas.) Interference Coefficient (cal.) Maintenance Select operating item Mode Sample Switch Gas Flow Time REF. Gas: Replacement Time REF. Gas: Measuring Time REF. Gas: Flowing Time SMP.
Page 70
Maintenance Set the gas key. Mode replacement time. Sample Switch 1 to 30 sec Gas Flow Time REF. Gas: Replacement Time REF. Gas: Measuring Time REF. Gas: Flowing Time SMP. Gas: Replacement Time SMP. Gas: Measuring Time SMP. Gas: Flowing Time Maintenance Select operating item Mode...
Page 72
6.7.1.3 Explanation of sample switch method Measuring time of Measuring time of Measuring time of Measuring time of reference gas sample gas reference gas sample gas Folw time of Folw time of Folw time of Folw time of reference gas sample gas reference gas sample gas...
6.8 Manual calibration procedure Measurement Mode Screen 6.8.1 Manual zero calibration ZERO ZERO Cal. Select Ch No. with UP / DOWN and ENT Back with ESC Range 1 0-50.00 ZERO Range 2 0-500.0 Range 1 0-50.00 Range 2 0-500.0 Range 1 0-10.00 vol% 0.00 Range 2 0-20.00...
Page 74
6.8.2 Manual span calibration measure- Measurement Mode Screen SPAN sensor. SPAN Cal. Select Ch No. with UP / DOWN and ENT Back with ESC SPAN Range 1 0-50.00 Range 2 0-500.0 Range 1 0-50.00 Range 2 0-500.0 Range 1 0-10.00 vol% 0.00 Range 2 0-20.00...
7.3 Long term maintenance Gas analyzer annual inspection plan levels in sample and reference gases). Year Recommended Component name Q’ty replacement Delivered 10th period (year) year year year year year year year year year year year Fuel cell O analyzer (build-in) Infrared light source O-ring for sampling cell Detector...
7.4 Cleaning of sampling cell CAUTION Maintenance actions should only be accomplished by properly trained and qualified personel. Not withstanding these maintenance steps, local facility and organizational safety program requrirements must be followed. 7.4.1 Disassembly and assembly of sampling cell a.
Page 78
Name Screw (for fixing the light source unit) Screw (for fixing the detector) Screw (for fixing the gas filter) Base plate Light source unit Screw (for fixing the support) Screw (for fixing the cell retainer) Gas filter Filter Support Cell retainer Pipe cell O-ring Infrared transmission window...
Page 79
b. How to remove block cell (See Fig. 7-2) Note) The O-ring (No. 9) is placed between the window holder and block cell. Take care about the O-ring position. With 2-component analyzer, install 2-component detector last. Take care so that no space is left between the 1-component and 2-component detectors.
Page 80
Name Screw (for fixing the light source unit) Filter Screw (for fixing the detector) Base plate Light source unit Screw (for fixing the block cell) Block cell Infrared transmission window (window holder) O-ring Screw (for fixing the measuring unit) Gas filter Detector Light source power PCB INZ-TN2ZPB-E...
Page 81
c. How to remove measuring unit (See Fig. 7-3) Note) Special care should be taken when assembling or disassembling the measuring cell to avoid the application of force to the detector pipe or light source unit pipe. If the pipe is deformed or damaged by excessive force, there is a danger of gas leak, thus resulting in misoperation.
7.4.2 How to clean sampling cell dow roughly. 7.5 Replacement of fuse Fuse holder Rear view Note) Prior to the following work, be sure to repair blown down fuse (short, etc), if any. INZ-TN2ZPB-E...
8. ERROR MESSAGE Error contents Error No.1 Light source/motor rotation is faulty. Error No.2 Error No.3 Error No.5 of full scale. Error No.6 Error No.7 50% of full scale. Error No.8 Error No.9 Error No.10 missed. INZ-TN2ZPB-E...
Page 84
Screen display and operation at the occurrence of error 0 0 8 Error No.9 0-50 1 3 6 0-50 0 0 0 0 0-20 vol% 0-50 2 1 0 0 0-25 vol% ZERO Cal. ENT : Go on calibration ZERO cal. error of selected CH.
nance mode. Error log screen Date and time when an error occurred. Maintenance ENT : Clear Error Log ESC : Back Error Log Component for which the error occurred. error No. No. 10 No. 9 Errors that occurred No. 5 No.
9. SPECIFICATIONS Analog input signal: For signal input from externally installed analyzer. Signal requirement; (1) Signal from Fuji’s Zirconia O analyzer Principle of measurement: NO, SO , CO , CO; (2) 0 to 1V DC from an O analyzer Non-dispersion infrared-ray absorption Input section is not isolated.
2. Standard Functions Output signal holding: Setting is variable within 1 to 99 hours Output signals are held unchanged during (in increments of 1 hour) or 1 to 40 days manual and auto calibrations by activation (in increments of 1 day). of holding (turning “ON”...
Page 88
4. Performance 21-On Repeatability: × Cs 21-Os Linearity: prior to atmospheric pressure correction (option) correction Zero drift: component of NDIR) sensor) concentration (Limit Span drift: Response time (T concentration 30 seconds or better (value changeable by setting.0 to Response interval may be changed de- pending on timing of the switching gas Average value after O correction and O...
Standard gas for calibration: Zero gas ; Dry N Span gas ; Each sample gas having its measuring range (recom- mended). In case a zirconia O analyzer is installed externally and calibration is carried out Zero gas ; Dry air or atmospheric air (Do not use with CO measure- ment)
Page 90
9.2 Measurable component and range - availability check table - Procedure of range selection On one component analyzer: First determine 1st range, then select 2nd range from the corresponding right column. More than two components analyzer: The 2nd range in the tables for two and more components is maximum available range. Select the 2nd range less than or equal to the “2nd range (max)”...
Page 91
2-component analyzer:NO/SO 2-component analyzer:NO/CO 1-component:NO 2-component:SO 1-component:NO 2-component:CO 1st range 2nd range (max) 1st range 2nd range (max) 1st range 2nd range (max) 1st range 2nd range (max) 0-50ppm 0-500ppm 0-50ppm 0-500ppm 0-50ppm 0-500ppm 0-50ppm 0-500ppm 0-100ppm 0-1000ppm 0-100ppm 0-1000ppm 0-100ppm 0-1000ppm 0-100ppm...
Page 92
3-component analyzer:NO/SO /CO >>> Combination of 1st component NO and 2nd component SO /3rd component CO 1-component:NO 1st range 2nd range (max) 0-50ppm 0-500ppm 0-100ppm 0-1000ppm 0-200ppm 0-2000ppm 0-250ppm 0-2500ppm 0-300ppm 0-2500ppm 0-500ppm 0-5000ppm 0-1000ppm 0-5000ppm 0-2000ppm 0-5000ppm 0-2500ppm 0-5000ppm 0-3000ppm 0-5000ppm 0-5000ppm...
Page 93
4-component analyzer:NO/SO /CO >>> 1st NO/4th CO and 2nd SO /3rd CO 1-component:NO 1st range 2nd range (max) 0-50ppm 0-500ppm 0-100ppm 0-1000ppm 0-200ppm 0-2000ppm 0-250ppm 0-2500ppm 0-300ppm 0-2500ppm 0-500ppm 0-5000ppm 0-1000ppm 0-5000ppm 0-2000ppm 0-5000ppm 0-2500ppm 0-5000ppm 0-3000ppm 0-5000ppm 0-5000ppm None 2-component:SO 4-component:CO 1st range...
9.4 Outline diagram (Unit : mm) Mounting method <TOP VIEW> The analyzer weight should be sup- ported at the bottom of the case. 19-inch rack mounting type <Rack dimensions> 450 or more Mounting diagram <FRONT VIEW> Guide rails Flowmeter Power Switch Guide rails <REAR VIEW>...
Page 98
www.fujielectric.com Instrumentation & Sensors Planning Dept. Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, japan Phone: +81-3-5435-7021 Fax: +81-3-5435-7475 www.fujielectric.com/products/instruments/...