10.5.19 Adjustment screen for analog output 100%

Description
 You can adjust the analog output 100% using this function.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
(6)
(7)
INZ-TN1ZKME-E
Adjusting the analog output 100% (20 mA)
Connect the ammeter to the analog output terminals (5) – (6).
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key.
The analog output adjustment screen appears.
Press the
key.
Press the
key.
The analog output 100% adjustment screen appears.
Adjust the analog output with the
Switch between "DOWN" and "UP" with the
Adjust the value to 20 mA with the
analog output with the ammeter.
Press the
key to set the value.
The display returns to the screen on the right.
Remove the ammeter connected to the analog output terminals
(5) – (6).
Description
key.
and
keys.
key, checking the
- 103 -
Displayed message
(LCD)
PARAMETER MENU
A-OUT ADJUST
A-OUT ADJUST
ANALOG OUT 0%
A-OUT ADJUST
ANALOG OUT 100%
ADJUST 
key.
100%
DOWN
A-OUT ADJUST
ANALOG OUT 0%