Station Number Setting - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

10.5.17 Station number setting

Description
 You can set the station number of the unit for MODBUS communication using this function.
 Settable range: 0 to 99
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
INZ-TN1ZKME-E
Setting the station number to 01
Display the screen on the right in accordance with the key
operation summary and press the
Use the
and
key to set the station number.
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
Description
key.
- 101 -
Displayed message
(LCD)
PARAMETER MENU
STATION NO
STATION NO
01
STATION NO
01
PARAMETER MENU
STATION NO