Ll Limit Setting Of Oxygen Concentration - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

10.5.7 LL limit setting of oxygen concentration

Description
 You can set the LL limit of oxygen concentration using this function.
Use the oxygen concentration alarm 1 for the range 1, and the oxygen concentration alarm 2 for the
range 2.
 Settable range: 0.001 to 55.000 vol%
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
(6)
Note
 A setting error occurs if the following condition is not satisfied:
"HH limit of oxygen concentration" ≥ "High limit of oxygen concentration" ≥ "Lower limit of oxygen
concentration" ≥ "LL limit of oxygen concentration"
Setting the Low-Low limit of oxygen concentration to "00.010 vol%" (Range 1)
Display the screen on the right in accordance with the key
operation summary and press the
Use the
key to select the oxygen concentration LL limit
value setting screen, and press the
Use the
and
key to set the oxygen concentration LL
limit value.
Press the
key to set the value.
Press the
key.
Press the
key.
The display returns to the screen on the right.
Description
key.
key.
- 90 -
Displayed message
(LCD)
PARAMETER MENU
OXYGEN ALARM 1
OXYGEN ALARM 1
L-LOW ALARM
L-LOW ALARM
00.010 vol%
L-LOW ALARM
00.010 vol%
OXYGEN ALARM 1
L-LOW ALARM
PARAMETER MENU
OXYGEN ALARM 1
INZ-TN1ZKME-E

Advertisement

Table of Contents
loading

Table of Contents