10.1.3 Full scale setting screen

Description
 You can set the full scale value for display of oxygen concentration value using this function.
 Settable range: If the decimal point position is set to "00.00": 02.00 to 50.00 vol%
If the decimal point position is set to "0.000": 2.000 to 9.000 vol%
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
INZ-TN1ZKME-E
Setting the full scale value to 20.00% (Range-1)
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key to display the screen on the right and press
the
key.
The full scale setting screen appears.
Use the
and
key to set the full scale value.
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
Description
key.
- 51 -
Displayed message
(LCD)
MEASURE MENU
OUTPUT RANGE
RANGE 1
FULL SCALE
FULL SCALE
20.00
FULL SCALE
20.00
RANGE 1
FULL SCALE