If a fuse blows, turn off the power switch, and replace the fuse after investigating the cause.
Open the front door and you can see the two fuses. The upper fuse is for protection of the converter circuit
and the lower one is for protection of the heater. Take care that these fuses are different each other in the
rated current.
To replace the fuse, insert a flathead screwdriver or coin into the fuse cap and turn it to the left while pressing
it in order to remove the cap and replace the fuse.
Put the cap on the fuse and turn it to the right to fix it.
For converter circuit
For heater
Note: Use time-lag fuses.
9.4 Replacement of fuse
or converter circuit
or heater
2.5A
The specifications of the fuse
Φ5×20 mm 0.5 A
(Example: 0213, 0.5 A, manufactured by Littelfuse)
Φ5×20 mm 2.5 A
(Example: 0213, 2.5 A, manufactured by Littelfuse)
- 46 -
0.5A
Specifications
INZ-TN1ZKME-E