Blowdown Menu (Option); Automatic Blowdown Setting - Fuji Electric ZKME Instruction Manual

Flameproof type zirconia oxygen analyzer converter
Table of Contents

Advertisement

10.3.1 Automatic blowdown setting

Description
 You can set the automatic blowdown to valid or invalid using this function.
 If changing the automatic blowdown setting from valid to invalid during automatic blowdown, the
blowdown is forcibly canceled.
Operation
(example)
Key operation
(1)
(2)
(3)
(4)
(5)
Note
 If the time for automatic calibration comes during manual treatments (calibration, blowdown, sensor
check, or sensor recovery) or remote treatments (calibration, blowdown, or heater off), the treatment
being performed is prioritized and the automatic calibration starts after the treatment is completed.
 If the time for automatic blowdown comes at the same time with automatic calibration, the automatic
blowdown starts first.
 If disconnection is detected (O
(combustion control: option)), or heater temperature error or A/D saturation error occurs, automatic
calibration is not performed.
INZ-TN1ZKME-E

10.3 Blowdown menu (option)

Setting the automatic blowdown to valid
Display the screen on the right in accordance with the key
operation summary and press the
Press the
key.
The auto blowdown valid/invalid setting screen appears.
Use the
key to select the auto blowdown valid (YES).
Press the
key to set the value.
Press the
key.
The display returns to the screen on the right.
sensor input, O
2
Description
key.
sensor thermocouple input, or thermocouple input
2
- 63 -
Displayed message
(LCD)
BLOW DOWN MENU
SET AUTO BLOW
SET AUTO BLOW
BLOW DOWN
BLOW DOWN
YES
NO
BLOW DOWN
YES
SET AUTO BLOW
BLOW DOWN