IRECTION
5-11-4-3
Input signals
Supply voltage of 5V and 12V from BEP power supply card.
The power voltages are regulated at +/-10%.
5V STDBY from the BEP power supply card for OPIO (Pass through).
PWR_ON signal from the Operator panel to the Motherboard (pass through).
Two USB ports for OP Panel (pass through), one USB for USB 'hub' and one USB for Patient IO
module.
Doppler Audio from GFI
Doppler audio mixed with Windows system sounds from motherboard
DVI (Digital Video) input from the video card or DVR Board.
AC Fail and PSON_N signal from the BEP PS.
5-11-4-4
Output signals
12V DC and 5V DC for Patient IO, DVD drive and High capacity drive.
5V STDBY from the BEP power supply card for OPIO (Pass through)
USB ports for Color and BW printer
3 x USB ports (Spare)
Audio Line In to the Motherboard
Audio to DVR board on Motherboard
Audio to the OP Panel for the speakers.
External Audio outputs
Audio to the sub woofer
DVI-D output for Main Monitor (Digital Video)
DVI-I output for External Monitor (only the Digital Video Signal is included)
Two external USB ports
Two spare internal USB ports
External Test connector
I
2
C bus clock and data signals to the Backplane and BEP PS.
5-11-5

Probe Connectors

The probe connectors are mounted on the Relay board.