Setup After Software Loading; Verifications After The Software Has Been Re-Loaded - GE VIVID E9 Service Manual

Hide thumbs Also See for VIVID E9:
Table of Contents

Advertisement

GE H
EALTHCARE
D
GA091568, R
IRECTION
8-4-10-7
Loading Software Patches
If a CD with a software patch is included, insert the disc in the drive, restart the VIVID E9 and follow the
on-screen instructions.
Please continue with
8-4-11

Setup after Software loading

NOTE:
Use this procedure if all partitions on the HDD have been erased during the software load.
NOTE:
If only the software on C:\ has been replaced or updated, please go to:
after the software has been re-loaded" on page
1.) Restore the Patient Archive and System Configurations from the backup you made before the
software loading.
-
2.) With your recordings from before the software loading, available, continue with the setup
instructions starting in:
3.) Continue with:
the settings as needed.
4.) Continue with:
settings as needed.
5.) After a software load, you should always calibrate the Front End, as described in:
Offset Calibration (Front End Alignment)" on page
6.) Install the network printer (if any). Please refer to the Printer Driver Installation Manual.
8-4-12

Verifications after the software has been re-loaded

NOTE:
DO NOT use this procedure if all partitions on the HDD have been erased during the software
load, but refer to:
NOTE:
Use this procedure if only the software on C:\ has been updated or reloaded.
1.) With your recordings from before the software loading available, continue with the setup instructions
starting in:
2.) Continue with:
the settings if needed.
3.) Continue with:
settings as needed.
4.) After a software load, you should always calibrate the Front End, as described in:
Offset Calibration (Front End Alignment)" on page
5
EVISION
8-4-12 "Verifications after the software has been re-loaded" on page
For instructions, please see "Data Backup and Restore" in the User Manual/User Guide.
Section 3-8 "Configuration" on page
Section 3-10 "Connectivity setup" on page
Section 3-11 "Options Setup" on page
8-4-11 "Setup after Software loading" on page
Section 3-8 "Configuration" on page
Section 3-10 "Connectivity setup" on page
Section 3-11 "Options Setup" on page
Chapter 8 - Replacement procedures
8-4-12 "Verifications
8-37.
3-27. Correct the settings as needed.
3-41. Based on your recordings, correct
3-49. Based on your recordings, correct the
6-11.
8-37.
3-27. Correct the settings if needed.
3-41. Based on your recordings, correct
3-49. Based on your recordings, correct the
6-11.
VIVID E9 S
M
ERVICE
ANUAL
8-37.
Section 6-6 "DC
Section 6-6 "DC
8 - 37

Advertisement

Table of Contents
loading

Table of Contents