3-7-4

Connections on the External IO

NOTE:
Accessory equipment connected to the analog and digital interfaces must be certified according
to the respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-
1 for medical equipment). Furthermore, all complete configurations shall comply with the valid
version of the system standard IEC60601-1-1. Everybody who connects additional equipment
to the signal input part or signal output part of VIVID E9, configures a medical system, and is
therefore responsible that the system complies with the requirements of the valid version of
IEC60601-1-1. If in doubt, consult the technical service department or your local representative
for GE Healthcare.
3-7-4-1
Connect Ethernet
Connect the Ethernet cable to the Ethernet connector on the External I/O (rear side of VIVID E9).
Figure 3-5 Ethernet connection on rear side of VIVID E9
3-7-4-2
Connect USB Flash Card
NOTE:
USB Flash Cards approved for VIVID E9 are verified for EMC performance according to
EN55011 class B. The use of any other USB Flash Cards will compromise this verification, and
may cause interference on VIVID E9 itself, or on other electronic devices.
For approved models, see:
Install the USB Flash Card in one of the USB ports on the VIVID E9. Two ports are provided on the
Operator Panel, to the left for the Touch Screen.
3 - 22
5
EVISION
L
R
Section 9-15 "Peripherals for VIVID E9" on page
Section 3-7 - Completing the setup
VIVID E9 S
!
ETHERNET
CONNECTOR
9-66.
M
ERVICE
ANUAL