GE VIVID E9 Service Manual page 439

Hide thumbs Also See for VIVID E9:
Table of Contents

Advertisement

GE H
EALTHCARE
D
GA091568, R
IRECTION
8-6-8-2
Remove the Frame w/LCD and Touch Screen
1.) Place the Operator Panel, Upper, with the front down on a clean, antistatic surface.
2.) Remove the three (3x) fixing screws (Phillips #1) and the two screws at the D-SUB connector (3/16
inch nut driver), see:
Figure 8-118 Operator Panel, Upper, seen from the rear side
NOTE:
On some VIVID E9s, the openings for the USB connectors have EMC fingers attached on the
right and left sides of each opening, ref.
when you remove the LCD Cover.
On newer VIVID E9s, the construction has been changed, so this issue has been resolved.
3.) Lift (pull) the upper part of the LCD Cover so you can release it from the four protuding tabs.
Figure 8-119 EMC fingers at openings for USB connectors
5
EVISION
Figure
8-118.
PULL TO REMOVE
Chapter 8 - Replacement procedures
PROTUDING TABS
(NOT VISIBLE IN THIS ILLUSTRATION)
Figure
8-119. Be careful to not loose these EMC fingers