Z-Mechanism removal procedure
The Z-Mechanism is fastened with several screws.
1.) Remove four (4x) screws from the lower part of the Z-Mechanism.
Figure 8-190 Remove four (4x) screws
2.) Remove four (4x) screws.
Figure 8-191 Remove four (4x) screws
3.) Lift the Z-Mechanism away.
Figure 8-192 Lift the Z-Mechanism away.