GE H
EALTHCARE
D
GA091568, R
IRECTION
8-7-4-2
Z-Mechanism removal procedure
The Z-Mechanism is fastened with several screws.
1.) Remove four (4x) screws from the lower part of the Z-Mechanism.
Figure 8-190 Remove four (4x) screws
2.) Remove four (4x) screws.
Figure 8-191 Remove four (4x) screws
3.) Lift the Z-Mechanism away.
Figure 8-192 Lift the Z-Mechanism away.
5
EVISION
Chapter 8 - Replacement procedures
VIVID E9 S
M
ERVICE
ANUAL
8 - 183