Initiate software reload from the Recovery Console
Ensure that the VIVID E9 is powered down.
1.) Depress the On/Off button on the Operator Panel. The VIVID E9 starts.
2.) Press the Esc button, on the alphanumeric keyboard, multiple times until the GRUB menu is
displayed on the screen.
Figure 8-2 GRUB menu
GNU GRUB version 0.97 (638K lower / 2095879K upper memory)
___________________________________________________________
Windows
Recovery Console
Use the | and | keys to select which entry is highlighted.
Press enter to boot the selected OS or 'p' to enter a
password to unlock the next set of features.
3.) On the Grub menu, select: Recovery Console.
The GE Healthcare Recovery Console is displayed.
Figure 8-3 GE Healthcare Recovery Console (Example)
4.) Call the Online Center (OLC).
Tell OLC that you need a "Challenge Code", and on request, you must read the Challenge String
on the screen. OLC will return a password that you need to type in as the Response.
5.) Enter the Response Code (password) you got from OLC.
8 - 14
5
EVISION
Section 8-4 - Reloading the software
VIVID E9 S
ERVICE
USE THE ARROW DOWN KEY
TO SELECT RECOVERY
CONSOLE, THEN PRESS
ENTER.
Enter the password (from OLC).